Is trabecular bone related to primary stability of miniscrews?
To compare the primary stability of miniscrews inserted into bone blocks of different bone mineral densities (BMDs) with and without cortical bone, and investigate whether some trabecular properties could influence primary stability. Fifty-two bone blocks were extracted from fresh bovine pelvic bone. Four groups were created based on bone type (iliac or pubic region) and presence or absence of cortical bone. Specimens were micro-computed tomography imaged to evaluate trabecular thickness, trabecular number, trabecular separation, bone volume density (BV/TV), BMD, and cortical thickness. Miniscrews 1.4 mm in diameter and 6 mm long were inserted into the bone blocks, and primary stability was evaluated by insertion torque (IT), mini-implant mobility (PTV), and pull-out strength (PS). Intergroup comparison showed lower levels of primary stability when the BMD of trabecular bone was lower and in the absence of cortical bone (P ≤ .05). The Pearson correlation test showed correlation between trabecular number, trabecular thickness, BV/TV, trabecular BMD, total BMD, and IT, PTV, and PS. There was correlation between cortical thickness and IT and PS (P ≤ .05). Cancellous bone plays an important role in primary stability of mini-implants in the presence or absence of cortical bone.ABSTRACT
Objective:
Materials and Methods:
Results:
Conclusion:
INTRODUCTION
The success of skeletal anchorage using miniscrews is related to their stability in the bone. Most of the failures occur immediately after mini-implant placement1 because the lack of primary stability may lead to progressive mobility of the device and to its subsequent loss.2 Factors that influence the immediate stability are related to the design of the device, the quantity and quality of bone, and the insertion technique.3
The term “bone quality” is not clearly defined in the literature. This includes physiological and structural aspects and the degree of bone tissue mineralization.4 Some bone properties, such as bone mineral density4–7 and cortical thickness,8–10 have been related to the stability of implants or mini-implants. Although the role of trabecular bone in the stability of dental implants11 has been investigated, there is still a lack of literature on miniscrews.
The aim of this study was to compare the primary stability of miniscrews inserted into bone blocks of different bone mineral densities (BMDs), as well as investigate whether there were any bone properties such as trabecular thickness, trabecular number, trabecular separation, cortical thickness, BMD, or bone volume density (BV/TV), that could influence the primary stability of miniscrews in the presence and absence of cortical bone.
MATERIALS AND METHODS
The sample consisted of 13 bovine pelvises (Bos taurus, Angus lineage) obtained from a slaughterhouse immediately after the animals were slaughtered. Four bone sections were taken from each pelvis—two from the pubic and two from the iliac region—with the use of a trephine bur (8 mm × 20 mm; Sin Implants, São Paulo, Brazil) adapted to a low-speed motor handpiece (Beltec LB100; Beltec, Araraquara, Brazil) under irrigation. In the samples taken from each bone region, the cortical bone was preserved in one and removed in the other using a diamond disc under irrigation. The final sample dimension was 8 mm × 10 mm. The material was immersed in saline solution and stored by freezing (−20°C).12 Division of the groups was based on thebone region and the presence of cortical bone (Table 1). The number of samples was calculated using the sample size data of a previous pilot study (α = 5%, power of study = 80%).

Evaluation of Bone Quality
Images of the samples were acquired in a micro-computed tomography (CT) system (Skyscan micro-CT, model 1173; Bruker, Kontich, Belgium) at a resolution of 9.3 µm using a 1-mm-thick aluminum filter, 80 kV, 90 µA, and exposure of 800 ms. The bone sections were kept in 2-ml Eppendorf tubes containing saline solution, with the bolt head facing upwards. The diameter of the Eppendorf was very close to that of the sample so that it was kept stable during image acquisition. The images were reconstructed (NRecon software; InstaRecon Inc, Ver 1 6.4 1, Champaign, Ill) and evaluated using the CT-Analyzer software (included with the Skyscan micro-CT).
The cortical thickness was measured in two-dimensional images. Sagittal and coronal sections were visualized in DataViewer (included with the Skyscan micro-CT), containing the center of the mini-implant. Two measurements were taken of each cross section, one on the left and one on the right side of the screw. The average of these four measurements was considered the cortical thickness. Thirty percent of the sample was measured twice in an interval of 1 week to assess the intraexaminer reliability by means of the intraclass correlation coefficient (ICC = .97).
In three-dimensional analysis, the volume of interest (VOI) corresponded to a cylinder 3.4 mm in diameter. The center of this cylinder, containing the screw, and the bone six voxels adjacent to it (54 µm) were excluded from the VOI in order to reduce the effect of artifact on analyses, a previously reported concern.12–15
In the imaging analysis process, a global threshold was used in order to distinguish trabecular bone (white pixels) from the background (empty space, black pixels) by means of a histogram analysis of grayscale images. Histomorphometric parameters such as trabecular thickness, trabecular number, and trabecular separation were automatically calculated for cancellous bone. For the total bone block, BV/TV was evaluated.
The BMD, in g/cm3 hydroxyapatite (HA), was computed from the attenuation values of grayscale in the micro-CT images. The phantom used to calibrate the software was an artificial bone block (containing 1.64 g/cm3 of 1-mm-thick cortical bone and 0.32 g/cm3 of trabecular bone; Sawbones, Vashon, Wash) with the same dimensions as those of the sample and containing a miniscrew inserted in its center. Thus, the conditions of the artifact were reproduced. The BMD was measured for the total bone block (total BMD), trabecular bone (trabecular BMD), and cortical bone (cortical BMD).
Insertion Torque
Fifty-two conical, self-drilling miniscrews (Ti-6Al-4V alloy; INP, São Paulo, Brazil) 1.4 mm in diameter and 6 mm long were inserted into the bone blocks. The implant sites were predrilled to a depth of 2 mm with a pilot drill 1.0 mm in diameter (INP). The mini-implants were placed by a single operator with the use of a manual key connected to a digital torque meter (Lutron TQ-8800; Lutron, Taipei, Taiwan). To guarantee perpendicular insertion of the miniscrews into the bone, the torque meter and the bone blocks were adapted to a mechanical device16 (Figure 1). The peak insertion torque (IT) values were recorded in Newton centimeters (Ncm).



Citation: The Angle Orthodontist 84, 3; 10.2319/052513-39.1
Miniscrew Mobility
Immediately after miniscrew placement, their mobility was evaluated with the Periotest instrument (Medizintechnik Gulden, Modautal, Germany). A special acrylic device was used to fix the sample and Periotest handpiece and to standardize the distance between the sleeve and the mini-implant (Figure 2). The handpiece was calibrated before each screw was measured. Two recordings were collected for each mini-implant, and the average value was designated as the Periotest value (PTV), which is on a scale from −8 to +50. A small PTV indicates low mobility and high primary stability.



Citation: The Angle Orthodontist 84, 3; 10.2319/052513-39.1
Pull-Out Strength
This mechanical test was performed in a universal test machine (Emic DL 2000; Emic. São José dos Pinhais, Brazil) using a 500 kgF load cell at a crosshead speed of 0.05 mm/s to remove the miniscrew16 (Figure 3). The maximum pull-out strength (PS) was recorded (N).



Citation: The Angle Orthodontist 84, 3; 10.2319/052513-39.1
Statistical Analysis
Data were evaluated with SPSS version 18 (SPSS Inc, Chicago, Ill). The normality and homogeneity of variables were verified by Shapiro-Wilk and Levene tests. Intergroup comparisons were performed by t-test (for cortical thickness and cortical BMD) and analysis of variance/Tukey tests (for the other variables). The Pearson correlation test was applied to verify the correlational relationships between variables. The level of significance was 5%.
RESULTS
Means, standard deviations, and intergroup comparisons for the bone properties are described in Tables 2 and 3. Trabecular thickness, trabecular number, trabecular separation, and trabecular BMD presented lower values in the iliac than in the pubic groups (Figure 4). BV/TV and total BMD presented increasing values from GI0 to GI1to GP0 to GP1 (Table 2). For GI1 and GP1, cortical thickness and cortical BMD did not differ (Table 3).



Citation: The Angle Orthodontist 84, 3; 10.2319/052513-39.1


Means, standard deviations, and intergroup comparisons for the variables used to evaluate miniscrew stability are shown in Table 4. IT values were lower for GI0, followed by GP0 and GI1 (without statistical difference between them), and the highest value was for GP1. PTV values in decreasing order were from GI0 to GI1 to GP0 to GP1. PS differed statistically among the groups, being higher in GP1, followed by GI1, then GP0, then GI0. Pearson correlation test results are presented in Table 5.


DISCUSSION
Practitioners have traditionally assessed primary stability by manual verification.17 Nevertheless, there are other less-subjective methods, such as IT; removal torque; PS, by which mechanical behavior can be assessed in the axial direction; and the mobility and lateral displacement test, by which mechanical behavior can be assessed laterally. However, there is still no gold standard for primary stability assessment.17 In this study, three methods were used for evaluating primary stability of mini-implants: IT, PTV, and PS.
The Periotest instrument allows a nondestructive and objective measurement of implant movement.18 It is probably the method that best reproduces the clinical verification of primary stability: manual verification of mobility. There is no consensus about the reliability of the Periotest for assessing implant stability. Some authors consider the Periotest a good tool for measuring stability in dental implants19–22 and mini-implants23,24; others disagree.25 In our study, there was correlation between the PTV value and the other two stability measures. IT presented a substantial negative correlation with PTV, as has previously been found.20,24,26 The correlation between PTV and PS was also negative. The two mechanical measures used to evaluate stability axially (IT and PS) presented strong positive correlation, in agreement with previous studies.16
It is known that a PTV of −8 to +9 corresponds to a mobility index of 0 with no distinguishable movement.27 In this study, the only group with PTV above +9 was GI0 (PTV = 20.19), which indicated palpable mobility.28
Considering bone properties, values for trabecular thickness, trabecular number, and trabecular BMD were higher for pubic bone, and trabecular separation was lower. These findings corroborate the difference in the trabecular characteristics of iliac and pubic bones.29 When the trabecular number increases, its separation diminishes and its thickness tends to increase. It reflects on BV/TV and BMD, which also increase. Both bone density measurements of the total block—BV/TV and BMD—showed very strong positive correlation. As previously mentioned,16 the BMD of trabecular bone and the total bone block presented a very strong positive correlation.
Iliac and pubic bone presented similar cortical bone characteristics. No difference was observed for cortical thickness and cortical BMD in GI1 and GP1. The presence of cortical bone per se numerically increased the primary stability, corroborating the importance of the cortical effect on primary stability.3,6,8–10,26 However, when the cortical bone was absent, the importance of the trabecular bone became more apparent. IT, PTV, and PS values differed between GI0 and GP0, with higher stability in pubic bone. When GI1 and GP0 were compared, IT and PTV values presented no statistical difference. It seems that the cortical bone plays an important role when trabecular bone has lower BMD, lower trabecular number, is thinner, and has more separated trabeculae. When comparing the stability values for GI1 and GP1, groups with similar cortical coverage, IT, PTV, and PS were higher for the pubic bone, which has a higher level of BMD, is thicker, and has less separate trabeculae and a higher trabecular number.
A positive correlation was found between cortical thickness and IT3,10,26,30–33 and between cortical thickness and PS.8,34 Cortical thickness did not influence the PTV value. It is believed that cortical thickness had a greater influence on axial than on lateral measures. Cortical BMD presented no valid correlation with IT, PS, or PTV, in disagreement with previous studies.16,24,16 However, the methodology of BMD evaluation differed. While this study evaluated BMD three-dimensionally by means of micro-CT, a more accurate method,35 the others used CT slices.
Although the results of the present study cannot be directly extrapolated to clinical practice because of the animal model and ex vivo methodology used, it was found that cancellous bone plays an important role in the primary stability of miniscrews, redirecting the spotlight that was previously focused only on the cortical bone.
CONCLUSIONS
-
As the BMD of the receptor site increases, the primary stability increases numerically as well;
-
In the presence and in the absence of cortical bone, trabecular number, trabecular thickness, BV/TV, trabecular BMD, and total BMD are correlated with the primary stability of mini-implants, showing the importance of trabecular bone in the stability of miniscrews.

Evaluation of IT. (a) Digital torque meter and bone block adapted to the device that was developed to allow insertion of mini-implants into the bone blocks perpendicular to the floor. (b) Approximate view of the process of mini-implant insertion.

Evaluation of PTV. (a) Periotest and sample adapted to the acrylic apparatus to maintain the tip of the Periotest handpiece parallel to the floor and perpendicular to the screw. (b) Close view of the Periotest tip maintained 2 mm away from the head of the mini-implant.

PS test. (a) Universal test machine during the test. (b) Close view of the devices manufactured to adapte the small sample to the machine: the lower one, used to fix the sample, and the upper one, used to extract the miniscrew.

Reconstruction of specimens from GI1 and GP1, respectively, showing the different architecture in trabecular bone.
Contributor Notes