Palatal volume and area assessment on digital casts generated from cone-beam computed tomography scans

,
,
,
,
, and
Page Range: 397 – 402
DOI: 10.2319/091117-611.1
Save
Download PDF

ABSTRACT

Objectives:

The objective of this study was to develop a reproducible method to measure the change of palatal volume and area through superimposition using maxillary expansion digital cast models.

Materials and Methods:

A total of 10 pre- and 10 postexpansion dental cast models were scanned by the same cone-beam computed tomography machine. Superimposition was performed using a fully automated surface-best fit of the palatal surfaces on the digital cast models. A gingival plane, identified only once on superimposed casts, and a distal plane with the lateral closing border and the palatal surface were used to localize this selection of air. Area and volume were calculated for pre- and postexpansion records. Pre- and postexpansion palatal volume and area were measured by the main investigator and three different observers for inter- and intra-observer reproducibility assessment.

Results:

The level of intra- and inter-observer agreement was very strong (intraclass correlation coefficients ≥ 0.953; P value < .0001) for all measurements.

Conclusions:

Palatal volume and area measurements based on the proposed superimposition are reproducible and can be used reliably.

INTRODUCTION

Palatal volume and area measurements on dental casts have always been challenging to accomplish even though those measurements can be useful in treatment evaluation because they express a more realistic assessment of clinical palate conditions. Palatal volume and area measurements can help evaluate changes from treatment modalities such as rapid palatal expansion1 and in the orthopedic treatment of cleft palate cases.2 Also, they can be used to show the relationship between anomalies in palatal morphology and function such as mouth breathing3 and posterior crossbite.4 In addition, they can be used for longitudinal evaluation of palatal vault anatomy.5 They usually require the use of dental casts, which are an integral part of orthodontic practice and research.6 Recently, the use of digital models derived from various scanning techniques has become more common. Having digital models allows clinicians to overcome some of the limitations of traditional plaster models as they are not subjected to physical damage and can be easily stored and transferred. Moreover, many studies have suggested that measurements derived from three-dimensional (3D) digital models have similar accuracy as those from traditional plaster casts.7 Optical scanners can be used both in vivo and in vitro to scan the dentition or plaster casts to create digital models.8

After making a 3D surface model of casts, the measurement of the palatal volume has been done geometrically or by creating a Computer-aided design model of the volume using reverse engineering technology.1,9 Alternatively, computed tomography and cone-beam computed tomography (CBCT) machines can provide volume data from which surface models can also be obtained, whereas laser and optical scanners can provide only mesh surface models that cannot be converted to volumetric data. The acquisition of 3D data through the CBCT scanning of dental casts and not directly from craniofacial structures is helpful because several factors complicate the analysis of in vivo scans. With in vivo scanning, there can be artifacts and distortion. Also, the palatal surface border is not easily detected, and the segmentation of the tongue is difficult due to its proximity to tissues such as the suprahyoid muscles and soft palate, which have similar Hounsfield units.10

Superimposition procedures can be used to assess treatment changes.11 Thus, CBCT digital models from casts can be used to provide a 3D superimposition analysis and to evaluate the changes after growth or treatment. Currently, there are no studies using the best-fit superimposition procedure to generate and calculate palatal area/volume changes. Thus, the aim of this study was to develop a reproducible method to measure changes in palatal volume and area after maxillary expansion through superimposition of digital cast models.

MATERIALS AND METHODS

A total of 10 pre- and 10 postexpansion dental cast models of patients from the Orthodontic Department, Faculty of Dentistry, Al Azhar University (Cairo, Egypt) were considered. The casts were randomly selected from consecutive patients, 5 females and 5 males, in the late mixed or early permanent dentition stage, with maxillary constriction associated with the presence of a uni- or bilateral posterior crossbite treated by rapid palatal expansion using a Haas expander appliance. The casts were not damaged and were free of gingival hypertrophy. The study was approved by the ethical committee of the same university (approval number REC16-032) in addition to University of Campania “Luigi Vanvitelli” (Naples, Italy; approval number 13933/17). In this study, the evaluation method was applied on expansion cases but it can be applied in any treatment modality.

Dental cast models were scanned by the same CBCT machine,12 with the same parameters, X-ray intensity, imaging time, and voxel size. SCANORA®3D (Soredex-Nahkelantie160, Tuusula, Finland) was used at 15 mA, 85 kV, and 20 seconds exposure time. DICOM-formatted images were generated with the same thickness, bit-depth, and dimensions, and then images of 0.35 thickness, 16 bit depth and 414 × 414 mm were produced. The Digital Imaging and Communications in Medicine (DICOM) files were then reconstructed to volumes to be imported into Computer-aided design software.

Several steps were performed for palatal volume measurement. First, CBCT scans were converted into a triangular mesh only to be used for superimposition.1315 A different color was assigned to each time record. Superimposition was performed using a fully automated surface-best fit of the whole area of pre- and postexpansion palatal surfaces after the exclusion of teeth.1,16,17 The two superimposed CBCT scans of digital models were then recorded as linked files and used as one (Figure 1). After this step, the mesh was no longer used, and all measurements were carried out on the linked superimposed volume data. Second, a gingival plane was constructed only on the pre-expansion record. The most gingival points on the lingual surface of canines, premolars, and the disto-lingual cusp of the first molars were identified. The software drew two imaginary lines intermediate between the canine, premolar, and molar gingival points, one for the right and another for the left side. Then the software made a projection from the canine gingival point and molar gingival point to these two lines. The four points constructed by the projection of the right and left canine and molar points on the two lines are shown in Figure 2. The gingival plane was constructed connecting the mid-canine point (a point midway between canine projections right and left) and the two molar projection points, right and left.

Figure 1. Superimposed pre-and postexpansion digital models based on best fit.
Figure 1 Superimposed pre-and postexpansion digital models based on best fit.

Citation: The Angle Orthodontist 88, 4; 10.2319/091117-611.1

Figure 2. Reference plane construction showing the two constructed imaginary lines.
Figure 2 Reference plane construction showing the two constructed imaginary lines.

Citation: The Angle Orthodontist 88, 4; 10.2319/091117-611.1

Three borders were constructed to localize the volume selection: gingival plane extended until its contact with the oral structures (Figure 3); distal plane, which was perpendicular to the gingival plane passing through a line connecting the two molar projections; and a lateral border. The remaining border was the palatal surface superior to the gingival plane and anterior to the distal plane. To construct a lateral border defining the volume and area, the digital model was reoriented on the gingival plane, and the gingival contour was traced manually and projected on the gingival plane (Figure 4).

Figure 3. Gingival plane and distal plane forming the occlusal and distal borders.
Figure 3 Gingival plane and distal plane forming the occlusal and distal borders.

Citation: The Angle Orthodontist 88, 4; 10.2319/091117-611.1

Figure 4. The lateral border to define the volume laterally.
Figure 4 The lateral border to define the volume laterally.

Citation: The Angle Orthodontist 88, 4; 10.2319/091117-611.1

On the superimposed digital models, the distal plane and lateral border were identified twice: once for pre- and once for postexpansion. The gingival plane was identified only once on the pre-expansion record.

Third, all air space was selected as a zero radio density and the model surface was selected, and then the two border planes, the lateral closing border, and palatal surface localized this selection of air and palatal surface inside only. Volume and area were then calculated for the pre- and postexpansion records.

The previous steps were used to explain to the software developers, but the actual steps performed by the operator were the following:

  • Identify gingival points manually.

  • Trace gingival borders.

  • Select zero radio density with the help of the software.

  • Localize volume selection with the help of the border planes.

All other construction was done automatically by the software, without operator intervention.

Reproducibility of Volume and Area Measurements

Pre- and postexpansion palatal volume and area were measured by the main investigator three times at weekly intervals and by three different observers once to evaluate inter and intraobserver reproducibility. The results were exported to an Excel spreadsheet. Means, standard deviations, coefficients of variation, and intraclass correlation coefficients for inter- and intraobserver reproducibility were evaluated.

RESULTS

Good precision of the method was evidenced by the narrow confidence interval. In addition, the sample size was adequate as demonstrated by the consistent range of the confident interval. The level of intra- and interobserver agreement was very strong (intraclass correlation coefficients ≥ 0.953; P < .0001) for all measurements (Tables 1 and 2). The coefficient of variation, which gives a measure of the variability of results regardless of the means, did not exceed 4.42% and was not correlated to palate size variation. It showed low variability (0.47–4.42%) across the range of the mean as demonstrated in Table 3.

Table 1 Intraclass Correlation Coefficient for Intraobserver Reproducibility Assessment of the Main Examiner

          Table 1
Table 2 Intraclass Correlation Coefficients for Interobserver Reproducibility Assessment

          Table 2
Table 3 Standard Deviation and Coefficient of Variation for Intra- and Interobserver Measures

          Table 3

DISCUSSION

Previously, to evaluate palatal area and volume, laser-scanned records have been used.1,9,18 All of the previously reported methods claimed accuracy and reproducibility of the border delimitation of the palatal space.2,19

In one method, the palatal space selection was based on delimited surfaces.35,18,20 A gingival plane was defined connecting the midpoints of the dento-gingival junctions. A distal plane was created through the two distal points of the second molars perpendicular to the gingival plane. These two planes were used for detecting the third border, such as the palatal surface. From these three boundaries, the palatal volume was calculated,35,18,20 but the palatal space was not selected directly as it is not technically possible on laser scan records. Furthermore, 10 3D points cannot be located geometrically in the same plane. Also, without a lateral border, it is not clear how the volume boundaries are determined in cleft cases or if some teeth are missing.

With another method, the selected volume was calculated by reverse engineering.1 The horizontal plane was produced connecting the following three identified points: one at the lowest point of the gingival margin of one central incisor and two at the lowest points of the gingival margin of the first permanent molars. The posterior limiting plane was tangent to the distal surface of the first molars and perpendicular to the horizontal plane. Palatal volume, visualized as a solid volume,1 was defined as the volume between the reference surfaces and the palatal surface. However, a lateral closing boundary was still missing. The gingival plane was produced from only three points. This may have led to loss of information. Moreover, it may not be comparable between consecutive records because two reference planes have been selected to measure pre- and postvolumes without reorienting the casts.2123

One additional method was based on geometric measurements performed by tracing and dividing the palatal space into small sections with regular shapes used to calculate area and volume geometrically.9 However, it followed straight lines and not the irregular surface, generating geometric shapes and lines that may not represent exactly the actual structural anatomy of the palate.

To overcome the limitations described, in this study a standardized method was used for the detection of a reference plane that was derived from three points constructed based on eight points. Using two imaginary lines instead of a best-fit plane may reduce the impact if there is an odd occlusal or gingival point, which may affect the whole plane. According to this reference plane, the digital cast was reoriented, allowing a reproducible gingival contour tracing to create a lateral border having good intra- and interclass correlations.

A scanned palatal surface does not have stable structure to be used for superimposition. Although palatal rugae have been suggested1 as relatively stable structures to use for superimposition, they have been shown to be hardly reliable as a reference.24 Consequently, the automated process of best-fit superimposition seems to have the least possible error.17,23

Moreover, using the same gingival plane for both pre- and postrecords can give more comparable results. Nevertheless, volume measurements calculated based on the superimposition of pre- and postdigital casts using the same reference plane has never been proposed. In this study, the gingival plane for superimposition was identified only once on the prerecord. The main advantage of this procedure is that the same operator or multiple operators can have reproducible and comparable results as shown by statistical analysis demonstrating no clinically significant method error. In addition, to avoid the influence of dental landmark changes resulting from orthodontic treatment such as tooth-supported expanders, unified border planes were created for both superimposed records to localize volume selection.

Other authors have assessed the reproducibility of palate volume measurement20 with a high variation of results, with standard deviations ranging between 34% and 40% of the mean and errors associated with palate size. In this study, the standard deviation did not exceed 9.7% of the mean and was nearly constant and not associated with palate size. The method in this study allowed good data reproducibility for both volume and area measurements in different patients. There was a low possibility of error in the identification of structures because the digital models were reoriented uniformly. Moreover, using the same gingival plane enabled easier comparison between pre-and posttreatment.

The data derived from CBCT exams can produce both volume and surface models, whereas that from laser scans can only be used to obtain surface models.17 Therefore, volume assessment does not require indirect calculations such as those produced geometrically or by reverse engineering.25 Instead, it simply follows from the direct selection of space inside the borders that have been selected. Higher resolution computed tomography machines can be used if better quality is needed.

This method has some limitations, such as the need for an available CBCT machine, although it could also be applied on a laser scan if suitable software tools were developed. It might take more time than other available methods, but it is reproducible. In the future, new software advances may decrease the time required.

CONCLUSIONS

  • Palatal volume and area measurements based on this proposed superimposition method were reproducible.

  • This novel approach represents a valid alternative to the other methods currently used to evaluate palatal volume and area.

REFERENCES

  • 1
    Gracco A.
    Malaguti A.
    Lombardo L.
    Mazzoli A.
    Raffaeli R.
    Palatal volume following rapid maxillary expansion in mixed dentition. Angle Orthod. 2010;80:153159.
  • 2
    Pisoni L.
    Codari M.
    De Menezes M.
    Ceron-Zapata AM.
    Lopez-Palacio AM.
    Dolci C.
    Longitudinal analysis of palatal volume in unilateral cleft lip and palate children. Italian J Anat Embryol. 2015;120(
    suppl
    ):79.
  • 3
    Lione R.
    Franchi L.
    Huanca Ghislanzoni LT, Primozic J, Buongiorno M, Cozza P. Palatal surface and volume in mouth-breathing subjects evaluated with three-dimensional analysis of digital dental casts-a controlled study. Eur J Orthod. 2015;37:101104.
  • 4
    Primožic J.
    Baccetti T.
    Franchi L.
    Richmond S.
    Farčnik F.
    Ovsenik M.
    Three-dimensional assessment of palatal change in a controlled study of unilateral posterior crossbite correction in the primary dentition. Eur J Orthod. 2013;35:199204.
  • 5
    Primožič J.
    Perinetti G.
    Richmond S.
    Ovsenik M.
    Three-dimensional longitudinal evaluation of palatal vault changes in growing subjects. Angle Orthod. 2012;82:632636.
  • 6
    Bell A.
    Ayoub AF.
    Siebert P.
    Assessment of the accuracy of a three-dimensional imaging system for archiving dental study models. J Orthod. 2003;30:219223.
  • 7
    Kasparova M.
    Grafova L.
    Dvorak P.
    et al.
    Possibility of reconstruction of dental plaster cast from 3D digital study models. Biomed Eng Online. 201331;
    12
    :49.
  • 8
    Hayashi K.
    Sachdeva AU.
    Saitoh S.
    Lee SP.
    Kubota T.
    Mizoguchi I.
    Assessment of the accuracy and reliability of new 3-dimensional scanning devices. Am J Orthod Dentofacial Orthop. 2013;144:619625.
  • 9
    Mikio K.
    Michiyo I.
    Tsutsu N.
    Eiki K.
    Harumi I.
    Michiharu D.
    Three dimensional measurements of the palate using a semiconductor laser: the influence of anterior cross bite on the palate in deciduous dentition. Pediatric Dental Journal. 2010;20:4044.
  • 10
    Shigeta Y.
    Ogawa T.
    Ando E.
    Clark GT.
    Enciso R.
    Influence of tongue/mandible volume ratio on oropharyngeal airway in Japanese male patients with obstructive sleep apnea. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;111:239243.
  • 11
    Nada RM.
    Maal TJ.
    Breuning KH.
    Bergé SJ.
    Mostafa YA.
    Kuijpers-Jagtman AM.
    Accuracy and reproducibility of voxel based superimposition of cone beam computed tomography models on the anterior cranial base and the zygomatic arches. PLoS One. 2011;6:e16520.
  • 12
    Berco M.
    Rigali PH Jr,
    Miner RM.
    DeLuca S.
    Anderson NK.
    Will LA.
    Accuracy and reliability of linear cephalometric measurements from cone-beam computed tomography scans of a dry human skull. Am J Orthod Dentofacial Orthop. 2009;136:17.e19.
  • 13
    Cevidanes LH.
    Motta A.
    Proffit WR.
    Ackerman JL.
    Styner M.
    Cranial base superimposition for 3-dimensional evaluation of soft-tissue changes. Am J Orthod Dentofacial Orthop. 2010;137:S120S129.
  • 14
    Cevidanes LH.
    Bailey LJ.
    Tucker GR Jr,
    Styner MA.
    Mol A.
    Phillips CL.
    et al.
    Superimposition of 3D cone-beam CT models of orthognathic surgery patients. Dentomaxillo Facial Radiol. 2005;34:369375.
  • 15
    Maes F.
    Collignon A.
    Vandermeulen D.
    Marchal G.
    Suetens P.
    Multimodality image registration by maximization of mutual information. IEEE Trans Med Imaging. 1997;16:187198.
  • 16
    Cha BK.
    Lee JY.
    Jost-Brinkmann PG.
    Yoshida N.
    Analysis of tooth movement in extraction cases using three-dimensional reverse engineering technology. Eur J Orthod. 2007;29:325331.
  • 17
    Choi J.
    Cha B.
    Brinkmann PG.
    Choi DS.
    Jang IS.
    Validity of palatal superimposition of 3-dimensional digital models in cases treated with rapid maxillary expansion and maxillary protraction headgear. Korean J Orthod. 2012;42:235241.
  • 18
    Primožič J.
    Franchi L.
    Perinetti G.
    Richmond S.
    Ovsenik M.
    Influence of sucking habits and breathing pattern on palatal constriction in unilateral posterior crossbite—a controlled study. Eur J Orthod. 2013;35:706712.
  • 19
    Veenema AC.
    Katsaros C.
    Boxum SC.
    Bronkhorst EM.
    Kuijpers-Jagtman AM.
    Index of Complexity, Outcome and Need scored on plaster and digital models. Eur J Orthod. 2009;31:281286.
  • 20
    Primožič J.
    Richmond S.
    Kau CH.
    Zhurov A.
    Ovsenik M.
    Three-dimensional evaluation of early crossbite correction: a longitudinal study. Eur J Orthod. 2013;35:713.
  • 21
    Gamba TO.
    Oliveira ML.
    Flores IL.
    Cruz AD.
    Almeida SM.
    Haiter-Neto F.
    Lopes SL.
    Influence of cone-beam computed tomography image artifacts on the determination of dental arch measurements. Angle Orthod. 2014;84:274278.
  • 22
    Alsufyani NA.
    Flores-Mir C.
    Major PW.
    Three-dimensional segmentation of the upper airway using cone beam CT: a systematic review. Dentomaxillofac Radiol. 2012;41:276284.
  • 23
    Frisardi G.
    Chessa G.
    Barone S.
    Paoli A.
    Razionale A.
    Frisardi F.
    Integration of 3D anatomical data obtained by CT imaging and 3D optical scanning for computer aided implant surgery. BMC Med Imaging. 2011;11:5.
  • 24
    Christou P.
    Kiliaridis S.
    Vertical growth-related changes in the positions of palatal rugae and maxillary incisors. Am J Orthod Dentofacial Orthop. 2008;133:8186.
  • 25
    Primozic J.
    Ovsenik M.
    Richmond S.
    Kau CH.
    Zhurov A.
    Early crossbite correction: a three-dimensional evaluation. Eur J Orthod. 2009;31:352256.
Copyright: © 2018 by The EH Angle Education and Research Foundation, Inc.
<bold>Figure 1</bold>
Figure 1

Superimposed pre-and postexpansion digital models based on best fit.


<bold>Figure 2</bold>
Figure 2

Reference plane construction showing the two constructed imaginary lines.


<bold>Figure 3</bold>
Figure 3

Gingival plane and distal plane forming the occlusal and distal borders.


<bold>Figure 4</bold>
Figure 4

The lateral border to define the volume laterally.


Contributor Notes

Corresponding Author: Professor Letizia Perillo, Head of Orthodontic Division and Chair of the Postgraduate Orthodontic Program, Department of Orthodontics, University of Campania Luigi Vanvitelli, Via L. De Crecchio 6, 80138, Naples, Italy (e-mail: letizia.perillo@unicampania.it)
Received: 01 Sept 2017
Accepted: 01 Jan 2018
  • Download PDF